Recirculating Cooler

13

Professional series

Temperature selection

digital

via keypad

indication on VFD COMFORT-DISPLAY

remote control via personal computer

indication on monitor

Temperature indication

°C

VFD COMFORT-DISPLAY

Resolution

°C

0.1

Temperature control

ICC - Intelligent Cascade Control


Electrical connections:

Computer interface RS232

External Pt100 sensor

Optional

Programmer input

-100 °C to 400 °C = 0 - 10 V or 0 - 20 mA or 4 - 20 mA

Input for the signal of a flow meter or external manipulated variable

Temperature recorder outputs

0 - 10 V (0 V = -100 °C, 10 V = 400 °C)

0 - 20 mA (0 mA = -100 °C, 20 mA = 400 °C)

4 - 20 mA (4 mA = -100 °C, 20 mA = 400 °C)

Standby input

for external emergency switch-off

Alarm output

for external alarm signal



Table 1

Pump capacity

Bath fluid: Water, Silicone -oil

XX.XX

PUMP

1

XX.XX

PUMP

2

XX.XX

PUMP
3

Circulating pump:

P0

P3 P0

P3 P0

Flow rate max.

Lpm at 0 bar

31

30 42

33 48

Pressure max.

bar at 0 liter

0.75

1.8 1.2

3.5 1.78

Bath fluid:

Galden

®

eg.

Fluorinert

®

3283

Pressure max.

bar at 0 liter

3.5 1,78

Notice:

If Galden

®

or Fluorinert

®

is used the the charge of the motor increases. Wrong

adjustment causes overheating and eventually destruction of the motor.
With the circulation pump P0 a maximum pump pressure stage >PUMP 2< may be
adjusted.
With the circulation pump P3 a maximum pump pressure stage >PUMP 1< may be
adjusted..
The pump type can be recognized in the order no. on the name plate.
95 xx xxx _ xx PX xx