Menu config - configuration – JULABO SC 10000w Professional Series Recirculating Coolers User Manual
Page 40
Menu functions
40
9.2.6.
Control parameters – XPU, XP, TN, TV external
Setting range: 0.1 ... 99.9
Setting range: 1 ...9999
Setting range: 0 ... 999
The control parameters preset in factory are in most cases adequate for
achieving an optimum temperature pattern for the samples requiring
temperature application.
Each parameter may be manually set via the keypad if necessary, to
allow optimum control performance.
1.
Press the
button until the desired submenu is displayed -
XP EXT, TN EXT, TV EXT, XPU EXT.
2.
Set value:
Use the cursor keys to move left or right on the display
until the numeral you wish to change blinks.
Use the increase/decrease arrows
to change the
selected numeral (-, 0, 1, 2, 3, ... 9).
3.
Press enter
to store the value.
Continue: Press
/
or
Setting range: 0.1 ... 99.9
Proportional range >Xpu<
The proportional range Xpu of the cascaded controller is only needed for
external control.
9.3. MENU CONFIG - configuration
CONFIG: Press enter
to switch to menu level 2
>SETP<
(KEY / SERIAL / PT100 / EPROG*)
Keypad control or remote control
>A-START<
(OFF / ON)
Autostart on / off
>OFFMODE<
(PMP OFF / PMP ON)
Motor on / off
>RESET<
( NO / YES ) Factory settings
>ACTVAR<
(CONTROL / SERIAL / EPROG* )
ACTuating
VARiable
>TIME<
(hh : mm ) Setting time
>DATE<
(TT/MM.JJ ) Setting date
EPROG* Indication only when electronic module is mounted.