9.2.6.

Control parameters – XPU, XP, TN, TV external


Setting range: 0.1 ... 99.9

Setting range: 1 ...9999

Setting range: 0 ... 999

The control parameters preset in factory are in most cases adequate for
achieving an optimum temperature pattern for the samples requiring
temperature application.
Each parameter may be manually set via the keypad if necessary, to
allow optimum control performance.

1.

Press the

button until the desired submenu is displayed -

XP EXT, TN EXT, TV EXT, XPU EXT.

2.

Set value:

Use the cursor keys to move left or right on the display

until the numeral you wish to change blinks.

Use the increase/decrease arrows

to change the

selected numeral (-, 0, 1, 2, 3, ... 9).

3.

Press enter

to store the value.

Continue: Press

/

or

Setting range: 0.1 ... 99.9

Proportional range >Xpu<

The proportional range Xpu of the cascaded controller is only needed for
external control.

9.3. MENU CONFIG - configuration

CONFIG: Press enter

to switch to menu level 2

>SETP<

(KEY / SERIAL / PT100 / EPROG*)

Keypad control or remote control

>A-START<

(OFF / ON)

Autostart on / off

>OFFMODE<

(PMP OFF / PMP ON)

Motor on / off

>RESET<

( NO / YES ) Factory settings

>ACTVAR<

(CONTROL / SERIAL / EPROG* )

ACTuating

VARiable

>TIME<

(hh : mm ) Setting time

>DATE<

(TT/MM.JJ ) Setting date


EPROG* Indication only when electronic module is mounted.