9.7. MENU ADJUST – ATC Absolute Temperature Calibration

ATC serves to compensate a temperature difference that might occur
between recirculating cooler and a defined measuring point in the bath
tank because of physical properties.

XXXXX

MENU
ADJUST

ADJUST: Press enter

to switch to menu level 2

>ATC SEN <

(INT)

or

(EXT)

> ATCSTAT <

> ATCSTAT <

(ON or OFF)

(ON or OFF)

>C ART

>C ART

>1- point <-, >2- point <- or >3- point < - calibration

> TTEMP 1<

> TTEMP 1<

> CTEMP 1<

> CTEMP 1<

> TTEMP 2<

> TTEMP 2<

> CTEMP 2<

> CTEMP 2<

> TTEMP 3<

> TTEMP 3<

> CTEMP 3<

> CTEMP 3<

1.

Press the

button until the desired submenu is displayed.

2.

Set value /parameter:

Use the cursor keys to move left or right on the display until
the numeral you wish to change blinks.

Use the increase/decrease arrows

to change the selected

numeral (-, 0, 1, 2, 3, ... 9).
or

select the parameter with the keys

.

3.

Press enter

to store the value / parameter

Continue: Press

/

or