Alarm code 12, Alarm code 14, Alarm code 15 – JULABO SC 10000w Professional Series Recirculating Coolers User Manual
Page 64: Alarm code 33, Alarm code 38, Warning code 40

Troubleshooting guide / Error messages
64
ALARM
CODE 12
Error in A/D converter
ALARM
CODE 14
Excess temperature sensors defect.
The excess temperature value lies below the working temperature
setpoint. Set the excess temperature to a higher value.
ALARM
CODE 15
External control selected, but external Pt100 sensor not connected or
defect.
ALARM
CODE 33
Cable of the excess temperature sensor interrupted or short-circuited.
ALARM
CODE 38
Ext. Pt100 sensor input without signal, but setpoint programming set to
external Pt100.
WARNING
CODE 40
The early warning system for low level signals a critical fluid level.
Replenish the bath tank with the bath fluid.
After eliminating the malfunction, press the mains power switch off and on
again to cancel the alarm state.
If the unit cannot be returned to operation, contact an authorized JULABO
service station.
Mains circuit breakers on the rear of the unit
4 Mains circuit breakers (resettable)16 A (with option H5)
2 Mains circuit breakers (resettable)10 A
0
1
Motor protection circuit breaker for compressor motor