Recirculating Cooler

69

Command

Parameter

Response of recirculating cooler

out_mode_08

1

Set the control dynamics - standard

out_sp_00

xxx.xx

Set working temperature. „SETP 1“

out_sp_01

xxx.xx

Set working temperature. „SETP 2“

out_sp_02

xxx.xx

Set working temperature. „SETP 3“

out_sp_03

xxx.xx

Set high temperature warning limit „OVERTEMP“

out_sp_04

xxx.xx

Set low temperature warning limit „SUBTEMP“

out_sp_06

xxx.xx

Set manipulated variable for the heater via serial interface
-99.99 ... +100 [%]

out_sp_07

x

Set the pump pressure stage. (1 ... 4)

out_par_04

x.x

CoSpeed 0 ... 5.0 Band limit during external control.
Setting the maximum difference between the temperatures
in the internal bath and external system.

out_par_06

xxx

Xp control parameter of the internal controller.

0.1 ... 99.9

out_par_07

xxx

Tn control parameter of the internal controller.

0 ... 9999

out_par_08

xxx

Tv control parameter of the internal controller.

0 ... 999

out_par_09

xxx

Xp control parameter of the cascade controller.

0.1 ... 99.9

out_par_10

xxx

Proportional portion of the cascade controller.

1 ... 99.9

out_par_11

xxx

Tn control parameter of the cascade controller.

0 ... 9999

out_par_12

xxx

Tv control parameter of the cascade controller.

0 ... 999

out_par_13

xxx

Maximum internal temperature of the cascade controller.

out_par_14

xxx

Minimum internal temperature of the cascade controller.

out_par_15

xxx

Band limit (upper)

0 ... 200 °C

out_par_16

xxx

Band limit (lower)

0 ... 200 °C