Recirculating Cooler
45
Warning:
The working temperature range of the recirculating cooler is determined during
configuration. If set to >CONTROL<, this range cannot be exceeded.
If set to > SERIAL < and > EPROG <, heat or cold is applied to the bath without
control. The permissible maximum temperature can be exceeded. The user has to
take adequate precautions for temperature control.
Materials, such as gaskets or insulations for example, may be damaged or
destroyed, if the permissible maximum temperature is exceeded.
The safety and warning functions of the instrument should always be used to their
fullest capacity.
8.
Safety installations, warning functions page 28
9.3.6.
TIME / DATE – setting time and date
XXXXX
TIME
16h4317
hh mm
XXXXX
DATE
10/1203
TT/MM.JJ
The integrated clock allows starting a profile at any date and time. The
clock is preset in the factory.
1.
Press the
button until the submenu > TIME < or > DATE < is
displayed.
2.
Setting time / date:
Use the cursor keys to move left or right on the display until
the numeral you wish to change blinks.
Use the increase/decrease arrows
to change the selected
numeral (-, 0, 1, 2, 3, ... 9).
3.
Press enter
to store the value.
Continue: Press
/
or
Clock: Only hours and minutes are set.
Settings are checked for plausibility.