39

9.2.5.

COSPEED - external

This parameter affects the temperature pattern only in case of external
control.

Possible parameters: 0.0 ... 5.0

1.

Press the

button until the submenu > COSPEED < is

displayed.

2.

Set value:

Use the cursor keys to move left or right on the display until
the numeral you wish to change blinks.

Use the increase/decrease arrows

to change the selected

numeral (-, 0, 1, 2, 3, ... 9).

3.

Press enter

to store the value.

Continue: Press

/

or

During selftuning, the control parameters Xp, Tn and Tv of a controlled
system are automatically determined and stored. Depending on the
controlled system, time for tuning can be unequally longer. This
controller layout allows protection of sensitive objects requiring
temperature application.

°C

t

Setpoint

5
3
0

Int. temperature

Ext. temperature

As soon as a co-speed factor is set, it is
considered for calculating the control
parameters. As shown in the diagram,
tuning times become shorter the higher the
co-speed factor is, but overshooting can
happen in the internal system.