beautypg.com

Menu limits, Set min -94.99 – JULABO SC 10000w Professional Series Recirculating Coolers User Manual

Page 47

background image

Recirculating Cooler

47

9.5. MENU

LIMITS

XXXXX

MENU
LIMITS

LIMITS: Press enter

to switch to menu level 2

>SET MAX<

>SET MIN<

>HEATMAX<

>COOLMAX<

>INT MAX<

>INT MIN<

>BAND H<

>BAND L<


In case of external control
these menu items are
additionally indicated.

1.

Press the

button until the desired submenu is displayed

2.

Set value:

Use the cursor keys to move left or right on the display until
the numeral you wish to change blinks.

Use the increase/decrease arrows

to change the selected

numeral (-, 0, 1, 2, 3, ... 9).

3.

Press enter

to store the value.

Continue: Press

/

or

Factory settings

XXXXX

SET MAX
300.00

XXXXX

SET MIN
-94.99

SETPOINT MAX / MIN -

Maximum and minimum setpoint

Restriction for the adjustable temperature range.

The limitation of the operating temperature range effects the temperature
setting under the menu called up with the

key.

It is possible to adjust only working temperatures that lie within the limit
range set here.
Existing settings for SETP 1, 2, 3 and also for >OVERTEMP< and
>SUBTEMP< (see page 30) are automatically defered within the limit
range.

Setting range:: -94.99 °C ... +300.0 °C