Recirculating Cooler

37

9.2.3.

DYN INT - Dynamic internal

Factory setting: APER (aperiodic)

°C

t

APER

NORM

Setpoint

temp. stability

°C

t

APER

NORM

temperature ramp

temp. stability

This parameter affects the temperature pattern only in
case of internal control.


Possible parameters:

NORM

Allows for reaching the setpoint faster, but

overshooting of up to 5 % is possible.

APER

Target temperature is attained without
overshooting.

With both adjustments an adequate temperature
stability is reached after approximately the same time.

1.

Press the

button until the submenu > DYN INT < is displayed.

2.

Select the parameter with the keys

.

(NORM / APER)

3.

Press enter

to store the selected parameter.

Continue: Press

or

9.2.4.

Control parameters – XP, TN, TV internal

The control parameters preset in factory are in most cases adequate for
achieving an optimum temperature pattern for the samples requiring
temperature application.
Each parameter may be manually set via the keypad if necessary, to
allow optimum control performance.

Setting range:0.1 ... 99.9

Proportional range >Xp<

The proportional range is the range below the selected temperature value
in which the control circuit reduces the heating power from 100 % to 0 %.

Setting range:1 ...9999

Resetting time >Tn< (Integral component)

Compensation of the remaining control deviation due to proportional
regulation. An insufficient resetting time may cause instabilities to occur.
Excessive resetting time will unnecessarily prolong compensation of the
control difference.