Recirculating Cooler

31

8.2.1.

Change-over of the warning function to shutdown function




Factory setting:
>WARNING<


For the two menu items > OVERTMP< and >SUBTMP< choose between a
warning message being signalled or a complete shutdown of the main
functional elements such as heater and circulating pump being effected
(see page 30).

 Setting >WARNING<

An audible warning (interval tone) sounds and a meassage appears on
the VFD COMFORT-DISPLAY.

XXXXX

WARNING
CODE 03

or

XXXXX

WARNING
CODE 04

OVERTMP SUBTMP

 Setting

>ALARM<

A complete shutdown of heater and circulating pump is effected.
An audible alarm (continuous tone) sounds and a message appears on
the VFD COMFORT-DISPLAY.

XXXXX

ALARM
CODE 03

oer

XXXXX

ALARM
CODE 04

OVERTMP SUBTMP

1.

Press the

button until the menu item >LIMITSR< is displayed.

2.

Select the parameter with the keys

.

(>WARNING< or >ALARM<)

3.

Press enter

to store the selected parameter.