JULABO SC 10000w Professional Series Recirculating Coolers User Manual
Page 31
Recirculating Cooler
31
8.2.1.
Change-over of the warning function to shutdown function
Factory setting:
>WARNING<
For the two menu items > OVERTMP< and >SUBTMP< choose between a
warning message being signalled or a complete shutdown of the main
functional elements such as heater and circulating pump being effected
(see page 30).
Setting >WARNING<
An audible warning (interval tone) sounds and a meassage appears on
the VFD COMFORT-DISPLAY.
XXXXX
WARNING
CODE 03
or
XXXXX
WARNING
CODE 04
OVERTMP SUBTMP
Setting
>ALARM<
A complete shutdown of heater and circulating pump is effected.
An audible alarm (continuous tone) sounds and a message appears on
the VFD COMFORT-DISPLAY.
XXXXX
ALARM
CODE 03
oer
XXXXX
ALARM
CODE 04
OVERTMP SUBTMP
1.
Press the
button until the menu item >LIMITSR< is displayed.
2.
Select the parameter with the keys
.
(>WARNING< or >ALARM<)
3.
Press enter
to store the selected parameter.