beautypg.com

Draining – JULABO SC 10000w Professional Series Recirculating Coolers User Manual

Page 25

background image

Recirculating Cooler

25

5.4. Draining

Notice:

Do not drain the bath fluid while it is hot or cold!

Check the temperature of the bath fluid prior to draining (by switching the unit on
for a short moment, for example).

Store and dispose of the used bath fluid according to the environmental

protection laws.

0

1

A

 Turn the mains switch (1) off.

 Place a suitable vessel for accepting the used bath liquid

underneath the drain.

 Unscrew the cup nut (A) from the drain port (25) and

empty the unit completely.

 Close the drain port.