Actvar control, Actvar serial, Actvar eprog – JULABO SC 10000w Professional Series Recirculating Coolers User Manual
Page 44
Menu functions
44
9.3.5.
ACTVAR - actuating variable
XXXXX
ACTVAR
CONTROL
Factory setting:
CONTROL
XXXXX
ACTVAR
SERIAL
XXXXX
ACTVAR
EPROG
The variable (ACTuating VARiable) corresponds to the extent to which
the heater or cooling machine of the recirculating cooler is controlled.
Heat or cold is applied to the bath according to this variable. If this
happens with the control electronics of the recirculating cooler, called
> CONTROL < in this particular case, the bath temperature is exactly
heated and maintained constant at the adjusted setpoint.
Programming of variables for the parameters > SERIAL < or >
EPROG < is only accepted, if the unit is in Start mode.
1.
Press the
button until the submenu > ACTVAR < is displayed.
2.
Select the parameter with the keys
(CONTROL / SERIAL / EPROG )
3.
Press enter
to store the selected parameter
Continue: Press
/
or
Possible parameters:
CONTROL
– The internal control electronics of the recirculating cooler
controls the heater and cooling machine. Self-tuning is possible.
SERIAL
– The heater or cooling machine receives the control signal via
the serial interface. Self-tuning is not possible.
EPROG -
The heater or cooling machine receives the control signal via
the E-Prog input. Self-tuning is not possible.
Important:
Under >MENU ANALOG< set the input variable to
>EPROG x / ACTVAR< (see page 61).
XXXXX
EPROG U
ACTVAR
Example: EPROG U / ACTVAR
Note:
The E-Prog input can only be used either under menu item
> SETP < (page 41) or under menu item > ACTVAR <.