JULABO SC 10000w Professional Series Recirculating Coolers User Manual

Page 60

background image

Menu functions

60

CHANNEL 1

CHANNEL 3

CHANNEL 2

REG+E-PROG

EPROG

1.

XXXXX

CHANNEL
3

1

REG+E-PROG :

Outputs of the connector REG+E-PROG

1.

Select CHANNEL 1, 2 or 3

2.

First define the desired output value for CHANNELs 1 to 3:
ACTINT

internal actual temperature value (bath temperature))

ACTEXT

external actual temperature value (external sensor)

POWER

periodic or intermittent heating or cooling

S-POINT

active setpoint temperature

(SETPoint 1, 2, 3,/ integr. programmer /external programmer)

2.

XXXXX

CH1
ACTINT


3.

XXXXX

CH1 0V
10.00

XXXXX

CH1 10V
210.00

3.

Then select the display size for CHANNELs 1 to 3:

Voltage outputs CHANNELs 1 and 2

Assign the voltage values of 0 V to the lowest and 10 V to the highest
necessary temperature (°C) or power rating (%) required as an output
value.

Current output CHANNEL 3

The current output (channel 3) offers 2 ranges for selection:
0 mA to 20 mA

or

4 mA to 20 mA

Assign the current values 0 mA or 4 mA to the lowest and 20 mA to
the highest temperature (°C) or power rating (%) required as an
output value.

XXXXX

CH3
ACTINT

XXXXX

CH3
0-20MA

XXXXX

CH3 0MA

xx.xx

XXXXX

CH3 20A

xx.xx

or

XXXXX

CH3
ACTINT

XXXXX

CH3
4-20MA

XXXXX

CH3 4MA
xx.xx

XXXXX

CH3 20A

xx.xx

(20A 20 MA)

0 V

0 mA

20 mA

4 mA

10 °C

210 °C

197 °C

202 °C

 5 K

 200 K

Example 1:

lowest temperature value:

10 °C

highest temperature value

210 °C

Fig.

shows

200

C scaled to paper width

rise: 50 mV/°C

Example 2:

lowest temperature value:

197 °C

highest temperature value:

202 °C

Fig.

shows

5

C scaled to paper width

rise: 2000 mV/°C