JULABO SC 10000w Professional Series Recirculating Coolers User Manual
Page 60
Menu functions
60
CHANNEL 1
CHANNEL 3
CHANNEL 2
REG+E-PROG
EPROG
1.
XXXXX
CHANNEL
3
1
REG+E-PROG :
Outputs of the connector REG+E-PROG
1.
Select CHANNEL 1, 2 or 3
2.
First define the desired output value for CHANNELs 1 to 3:
ACTINT
internal actual temperature value (bath temperature))
ACTEXT
external actual temperature value (external sensor)
POWER
periodic or intermittent heating or cooling
S-POINT
active setpoint temperature
(SETPoint 1, 2, 3,/ integr. programmer /external programmer)
2.
XXXXX
CH1
ACTINT
3.
XXXXX
CH1 0V
10.00
XXXXX
CH1 10V
210.00
3.
Then select the display size for CHANNELs 1 to 3:
Voltage outputs CHANNELs 1 and 2
Assign the voltage values of 0 V to the lowest and 10 V to the highest
necessary temperature (°C) or power rating (%) required as an output
value.
Current output CHANNEL 3
The current output (channel 3) offers 2 ranges for selection:
0 mA to 20 mA
or
4 mA to 20 mA
Assign the current values 0 mA or 4 mA to the lowest and 20 mA to
the highest temperature (°C) or power rating (%) required as an
output value.
XXXXX
CH3
ACTINT
XXXXX
CH3
0-20MA
XXXXX
CH3 0MA
xx.xx
XXXXX
CH3 20A
xx.xx
or
XXXXX
CH3
ACTINT
XXXXX
CH3
4-20MA
XXXXX
CH3 4MA
xx.xx
XXXXX
CH3 20A
xx.xx
(20A 20 MA)
0 V
0 mA
20 mA
4 mA
10 °C
210 °C
197 °C
202 °C
5 K
200 K
Example 1:
lowest temperature value:
10 °C
highest temperature value
210 °C
Fig.
shows
200
C scaled to paper width
rise: 50 mV/°C
Example 2:
lowest temperature value:
197 °C
highest temperature value:
202 °C
Fig.
shows
5
C scaled to paper width
rise: 2000 mV/°C