Safety installations, warning functions

28

8.

Safety installations, warning functions

Check the safety installation at least twice a year! See page 21

SAFETMP

OVERTMP

SUBTMP

LIMITSR

Settings for the excess temperature protection > SAFETMP< according to
IEC 61010-2-010 and for the high > OVERTMP< and low> SUBTMP<
temperature warning functions are made in a menu that is called up with

the key

.


Menu item > LIMITSR <: „Warning“ or „Alarm“
For the two menu items > OVERTMP< and >SUBTMP< choose between
a warning message being signalled or a complete shutdown of the main
functional elements such as heater and circulating pump being effected.

8.1. Excess temperature protection

XXXXX

ALARM
CODE 1

°C

°C

160

240

320

0

80

40

XXX.X

SAFETMP
100

This safety installation is independent of the control circuit. When the
temperature of the bath fluid has reached the safety temperature, a
complete shutdown of the heater and pump is effected.

The alarm is indicated by optical and audible signals (continuous tone)
and on the VFD COMFORT-DISPLAY appears the error message
"ALARM-CODE 1".

Setting range: 20 °C ... 220 °C

1.

Press the

button until the menu item > SAFETMP < is

displayed.

2.

Set the new cut-out value using a screwdriver via the VFD
COMFORT-DISPLAY (Example: 100 °C)

3.

Press

to update the display immediately, or the unit

automatically returns to the effective display after

about 30 seconds

.

Recommendation:

Set the excess temperature protection at 5 to 10 °C above the working
temperature setpoint.

Warning:

The excess temperature protection >SafeTemp< should be set at least 25 °C below
the fire point of the bath fluid used.
There is a risk of fire in the event of a wrong setting!
We disclaim all liability for damage caused by wrong settings!