Technical specifications – JULABO SC 10000w Professional Series Recirculating Coolers User Manual
Page 11

Recirculating Cooler
11
2.4. Technical
specifications
Professional Series
SC 5000a
SC 5000w
Working temperature ranges:
Standard unit
°C
+5 ... +35
+5 ... +35
Option Low Temp
°C
-20 ... +35
-20 ... +35
Option Low /HighTemp I
°C
-20 ... +80
-20 ... +80
Option Low /HighTemp II
°C
-20 ... +130
-20 ... +130
Temperature stability
°C
0.1
0.1
A
bsolute Temperature Calibration INT/EXT
±3 / ±9
±3 / ±9
Heater wattage (at 230 V) Option H5
kW
5.0
5.0
Heater wattage (at 230 V) Option H1
kW
12.0
12.0
Cooling capacity
°C
+20 0 -10
+20 0 -10
Medium ethanol
kW
5.0 2.5 1.2
5.0 2.5 1.2
Refrigerant R404A
R404A
Pump capacity P3 (Standard) / P4 * (Option)
see table 1 page 13
Flow rate
P3 / P4
l/min at 0 bar
33 / 43
33 / 43
Pressure max.
P3 / P4
bar at 0 liters
3.5 / 4.4
3.5 / 4.4
Overall dimensions (WxDxH)
cm
59x67/112
59x67/112
Filling volume
liters
43 ... 60
43 ... 60
Weight
kg 153
153
Ambient temperature
°C
5 ... 40
5 ... 40
Mains power connection
V/ Hz
365-440 V/3PNPE/50 Hz
400/50/3 Phases
400/50/3 Phases
Current consumption without heater
A
11 (at 400 V)
11 (at 400 V)
Current consumption 5 kW heater
A
22/Phase
22/Phase
Current consumption 12 kW heater
A
27/Phase
27/Phase
Mains power connection
V/ Hz
192-254 V/3PPE/60 Hz
208-230/60/3 Phases
208-230/60/3 Phases
Current consumption without heater
A
19/Phase (at 208 V)
--------
Current consumption 5 kW heater
A
17/ Phase (at 208 V)
17/Phase (at 208 V)
Current consumption 12 kW heater
A
36/Phase (at 230 V)
36/Phase (at 230 V)
All measurements have been carried out at: rated voltage and frequency ambient temperature: 20 °C
Technical changes without prior notification reserved.
* Pump P4- reduces cooling capacity by 0.3 kW