Menu control – JULABO SC 10000w Professional Series Recirculating Coolers User Manual
Page 55

Recirculating Cooler
55
Example:
1-point calibration
°C
°C
T 1
T
3-point calibration
°C
°C
T 1
T
T
T
1 = Original curve
Principle
:
For ATC calibration, in steady state the bath temperature at the location
of the temperature sensor (CT) is determined at the respective adjusted
working temperature. This value is then set on the recirculating cooler in
the menu >ATCalibration< under menu item > CTEMP X <.
This can be a 1-point, 2-point or 3-point calibration.
M
CT
B
T
TT
M = Temperature measuring instrument with temperature sensor
B = Bath tank with T = recirculating cooler
CT = Temperature on measuring point
TT = Temperature on recirculating cooler
Example:
3-point calibration for internal control.
In the temperature range of 80 °C to 160 °C the calibration curve of the
temperature sensor (TT) should be assimilated to the actual
temperatures on the measuring point (CT).
XXXXX
MENU
CONTROL
Set controller to internal control:
The control type can only be adjusted in the -OFF- condition
1.
Press the
button until the menu item > MENU / CONTROL< is
displayed and press enter
.
1.1. Under the menu >CONTROL< set the parameter to > INT < and
press enter
(see page 34).
Continue: Press
.
2.
Press the start/stop key
.