Menu control – JULABO SC 10000w Professional Series Recirculating Coolers User Manual

Page 55

background image

Recirculating Cooler

55

Example:
1-point calibration
°C

°C

T 1

T


3-point calibration
°C

°C

T 1

T

T

T

1 = Original curve

Principle

:

For ATC calibration, in steady state the bath temperature at the location
of the temperature sensor (CT) is determined at the respective adjusted
working temperature. This value is then set on the recirculating cooler in
the menu >ATCalibration< under menu item > CTEMP X <.
This can be a 1-point, 2-point or 3-point calibration.

M

CT

B

T

TT

M = Temperature measuring instrument with temperature sensor

B = Bath tank with T = recirculating cooler

CT = Temperature on measuring point
TT = Temperature on recirculating cooler





Example:

3-point calibration for internal control.
In the temperature range of 80 °C to 160 °C the calibration curve of the
temperature sensor (TT) should be assimilated to the actual
temperatures on the measuring point (CT).

XXXXX

MENU
CONTROL

Set controller to internal control:

The control type can only be adjusted in the -OFF- condition

1.

Press the

button until the menu item > MENU / CONTROL< is

displayed and press enter

.

1.1. Under the menu >CONTROL< set the parameter to > INT < and

press enter

(see page 34).

Continue: Press

.

2.

Press the start/stop key

.