JULABO Magnum 91 Temperature System User Manual
Page 72

Remote control
72
Error messages
Description
-46 COMPRESSOR STAGE 2
PRESSURE TO HIGH FOR A LONG 
TIME ON LOW PRESSURE SIDE 
The compressor of stage 2 does not start. 
5 minutes after stage 1 is switched on, the pressure of stage 2 
does not fall below a determined value. Therefore stage 2 does 
not start. 
Cancel the alarm state by turning the mains switch off (wait about 
4 seconds) and by turning the switch on again in order to restart 
the unit.
If the error reappears after switching the unit on again, the 
refrigerant circuit of stage 1 needs to be checked by a specialist. 
 
-47 COMPRESSOR STAGE 2 TO LONG
ABOVE HIGH PRESSURE
The cooling system is surcharged, and the compressor of stage 2 
was switched off. 
Before the WARNING CODE 42 is shown for more than two 
hours. In order to avoid the destruction of the compressor, the 
refrigerant cycle of stage 2 should be checked by a specialist. 
 
-48 COMPRESSOR STAGE 2 TO OFTEN
SWITCH ON/OFF
Caused by WARNING CODE 43 the compressor of stage 2 has 
switched off 5 times within one hour. 
If after a certain time there was a pressure compensation, the 
alarm status can be cancelled by switching off and on at the mains 
switch, and the unit is ready to operate again. 
In order to avoid the destruction of the compressor, the refrigerant 
circuit should first be checked by a specialist. 
 
-49 PRESSURE SENSOR ALARM LOW
PRESSURE SIDE STAGE 2
The pressure sensor of stage 2 (evaporation pressure, low 
pressure side) is bypassed or has a line interruption. 
Have the repair done by a specialist. 
 
-50 PRESSURE SENSOR ALARM HIGH
PRESSURE SIDE STAGE 2
The pressure sensor of stage 2 (high pressure side) is bypassed 
or has a line interruption. 
Have the repair done by a specialist. 