JULABO Magnum 91 Temperature System User Manual
Page 28

Preparations
28
5.6.1.
Filling of external, closed systems
Filling is performed in two sections. In section 2, the temperature system 
starts an automatic filling process enabling convenient filling. 
Notice:
First connect the external system! (see page 23)
-OFF-
ALARM
CODE 01
Section 1
 Connect the unit to a mains power socket (see page 26) and turn on
the unit with the mains power switch (1). 
During the self-test all segments of the VFD-Info-Display, all control 
indicators and the DIALOG-DISPLAY light up. 
After the self-test, the VFD-Info-Display signals low liquid level alarm. 
>CODE 01< and a signal tone sounds. 
Press enter
to quit the audible signal.
 Using a filling funnel, slowly fill in the bath fluid via the filling opening
(5) until the level indication
indicates 3 or 4 segments. Watch the
level indication
.
Section 2
5
REG+E-PROG
STAND-BY
ALARM
SERIAL
ESC
Soll 1: xxx.xx°C
IntIst: xxx.xx°C
Leistung: 50%
Regelung: Intern
°C
10
30
60
130
180
220
°C
0
40
80
160
240
320
EXT
Pt100
FILL
-OFF-
 AUTO
 FILL
 Turn the unit off (approx. 4 seconds) and on again with the mains
power switch (1). Wait until the self-test is completed.
 Press the key
to start the automatic filling process. On the
VFD-Info-Display „AUTO FILL“ is shown, in intervals. 
As soon as the pump pressure indication in the 3
rd
line of the VFD-
Info-Display 
(„P X.XX“) has scaled up to at least 0.1 bar, the tempering liquid is 
pumped into the external system. In intervals of 10 seconds bath 
liquid is pumped into the external system - air purge is carried out in 
the breaks. 
Notice: The filling mode has to stay activated for at least 5 minutes 
in order to fully de-air the system. 
-OFF-
Warning
CODE 40
 The liquid level in the unit falls.
In case of a WARNING >CODE 40< again slowly fill in bath fluid into 
the opening (5). 
Watch the level indication 
.
In case of a ALARM >CODE 01< return to the final condition of 
section 1 and restart section 2. 
 If the liquid level display remains unchanged after some minutes,
complete filling according to the application task. 
See page 27 - marker "HT" or "TT". 
Pump
>Pump OFF
Modu: Sys close
 Press
to finish the automatic filling process.
 In the submenu >Pump< the unit has to be set on
>Mode:< >sys close< . (see page 29)
 The unit is now ready for operation .
 
 
