JULABO Magnum 91 Temperature System User Manual
Page 47
Temperature System Magnum 91
47
8.3.1. Interrupting
a
profile
Setp. : xx.xx°C
IntAct: xx.xx°C
Prof. : x Stp: x
Remain: xx:xx:xx
*** pause ***
Interrupting a profile:
Press the start/stop key
to interrupt or continue a profile.
The setpoint and time period set for the corresponding section are
thus stopped at the values presently achieved.
The instrument is put on hold and the message "pause" flashes on
the DIALOG DISPLAY (LCD).
AK
1
2
3
STAND-BY
Setp. : xx.xx°C
IntAct: xx.xx°C
Prof. : x Stp: x
** STAND-BY **
STAND-BY
A profile can be interrupted or restarted by an external emergency
shut-off (see page 63).
CAUTION:
This is not an actual emergency switch-off.
The setpoint control and the timer are interrupted by breaking the
contact "AK”.
The instrument is put on hold and the message "Stand-By" flashes on
the DIALOG-DISPLAY (LCD).
Important:
To achieve this, the stand-by condition must first be activated and the
AUTOSTART function turned on
(See page 42).
Warning:
Following a power interruption, it would be possible in this condition for the instrument to
restart automatically. The safety and warning functions of the instrument should always be
used to their fullest capacity.
See Warning page 42.
>Start Profile 0
at Step 0
Loops 1
Start yes
Setp. : xx.xx°C
IntAct: xx.xx°C
Prof. : x Stp: x
Remain: xx:xx:xx
Termination of a profile:
A profile can be terminated by pressing the escape key
ESC
. The
programmer switches back to the Start menu.
Press the escape key
ESC
again to leave the menu
or
use the cursor keys
to remain in the Start menu.
The execution of another temperature profile can now be prepared if
necessary.