Operating procedures

32

6. Operating

procedures

6.1. Switching on / Selecting the language

JULABO
Magnum
91
V1.91-AE

Switching on:

IN

The unit is operated by turning the mains power switch (1).

The integrated pilot lamp illuminates.

During the self-test all segments of the VFD-Info-Display, all control
indicators and the DIALOG-DISPLAY light up.
Then the software version number (example: V1.91-AE) appears for
a short moment and the message "OFF" indicates the unit is ready to
operate.

Message >WAIT< when adjusting >Mode: sys close<:.
During the message >WAIT< the ventilation valve is opened for 10
seconds. So, the air which has eventually been collected during the
standstill, can escape.

The unit returns to the previous operating mode that was active
before it was turned off:
keypad control mode

(manual operation via the unit itself/RD)

or
remote control mode

(operation via PC).

Selecting the language:
There are two options for the language of the DIALOG-DISPLAY (LCD):
German or English. Select the desired language in the MENU level under
the submenu >Configuration.

Soll: 20.00°C
IntIst xxx.00°C
ExtIst:
Regelung:

>Konfiguration
Regelparameter

Konfiguration
Identif. xxxxx
Sollwert xxxxx
Autostart xxx
Standby xxxx
>Sprache xxxxxxx

Setp.: 20.00°C
IntAct xxx.xx°C
ExtAct: xxx.xx°C
Control: intern

Press the respective keys in the following order:

1. MENU key

1x

2. Enter key

1x

3. Cursor key

up to Submenu "Sprache/Language"

4. P-key

1x

5. Enter key

1x

6. Escape key

ESC

2x


The DIALOG-DISPLAY (LCD) helps to follow up the individual settings
(example: swap the language from German to English).