JULABO Magnum 91 Temperature System User Manual
Page 70

Remote control
70
Error messages
Description
-13 WARNING : VALUE EXCEEDS
TEMPERATURE LIMITS 
 
Value lies outside the adjusted range for the high and low 
temperature warning limits. But value is stored. 
-14 TEMPERATURE/LEVEL ALARM
Safety temperature alarm
-15 EXTERNAL SENSOR ALARM
External control selected, but external Pt100 sensor not 
connected. 
-16 TRIAC/RELAY CONNECTION
OPEN 
 
Heating circuit interrupted. 
 
-17 TRIAC SHORTED
Heating circuit short-circuited. 
 
-20 WARNING: CLEAN
CONDENSOR OR CHECK 
COOLING WATER CIRCUIT OF 
REFRIGERATOR 
 
Cooling of the condenser is affected. 
Clean air-cooled condenser. 
Check the flow rate and cooling water temperature on water-
cooled condenser. 
-21 WARNING: COMPRESSOR
STAGE 1 DOES NOT WORK
Compressor stage 1 does not work. 
 
 
-26 WARNING: STAND-BY PLUG IS
MISSING 
 
External stand-by contact is open. 
Stand-by input - see pages 42 and 63. 
-31 INTERNAL COMMUNICATION
ERROR 
 
Communication between remote module RD and instrument 
interrupted. 
-40 NIVEAU LEVEL WARNING
Low liquid level warning in the internal reservoir.
-27 WARNING:
COMPRESSOR STAGE 1 
OVERPRESSURE OR OVERTEMP 
 
The compressors of both stages were switched off. 
 The high pressure controller in stage 1 measures an 
excessive pressure. As soon as the pressure in stage 2 
reaches the lower switching point, the compressors are 
switched on again one after the other. 
 The motor winding protection in stage 1 measures an
excessive temperature. 
As soon as the temperature in stage 1 reaches the lower 
switching point, the compressors are switched on again one 
after the other. 
Check the cooling water supply line. 
Wait until the pressure compensation in stage 1 is finished 
and switch on the unit again after waiting for about 25 
minutes. 
-28 WARNING:
COMPRESSOR STAGE 1 HOTGAS-
OVERTEMP 
The compressor of both stages were switched off. 
The exit temperature of stage 1 is too high. As soon as the 
temperature reaches the lower switching point, and – as second 
condition – the pressure proportions allow it, the compressors are 
switched on again delayed. 
 