beautypg.com

JULABO Magnum 91 Temperature System User Manual

Page 17

background image

Temperature System Magnum 91

17

3.7

°C

10

30

60

130

180

220

Adjustable safety temperature.
Used for setting the safety temperature in the internal reservoir, called "RES"
on the display

3.8

Key for automatic filling and air purge of the unit as well as the connected
external system.

4 Socket

board

4.1

REG+E-PROG

Programmer input and temperature recorder output

4.2

STAND-BY

Stand-by input (for external emergency switch-off)

4.3

ALARM

Alarm output (for external alarm signal)

4.4

EXT. Pt100

Connector for external measurement and control sensor

4.5

SERIAL

Interface RS232/RS485

5

Filling

opening

6

Drain port with drain tap

7a /
7b

Castor with brake (at the front)
Castor without brake (at the back)

8

F1 F2 F3 F4 F5 F6

Mains power connection 400 V / 3 Phases / 50 Hz
Mains fuses
F1 and F4

Mains circuit breakers (resettable): 10 A Pump motor

F2 and F5

Mains circuit breakers (resettable): 16 A Heater 1

F3 and F6

Mains circuit breakers (resettable): 16 A Heater 2, Electronics

8

F1 F2 F3 F4 F5 F6

Mains power connection 230 V / 3 Phases / 60 Hz
Mains fuses
F1 and F2

Mains circuit breakers (resettable): 16 A Heater 2, Electronics

F3 and F4

Mains circuit breakers (resettable): 16 A Heater 1

F5 and F6

Mains circuit breakers (resettable): 10 A Pump motor

8

0

1

0

1


2 Motor protection circuit breakers for compressor motors

9

Mains

power

cable

10

Pump connector: pressure pump

M24x1.5

11

Pump connector: return

M24x1.5

12

Overflow connector M16x1

13
14

Cooling water connectors IN - inlet

OUT - outlet

G3/4" external thread