9. Troubleshooting guide / Error messages

XXXXX

ALARM
CODE 03

+

Alarm with a complete shutdown of the unit
Whenever the microprocessor electronics registers a failure, a complete shutdown of

the heater and circulating pump is performed. The alarm light „

“ illuminates and a

continuous signal tone sounds.
The VFD-COMFORT-DISPLAY indicates the cause for the alarm in form of a code.

XXXXX

WARNING
CODE 3

Warning without a complete shutdown of the unit
The VFD-COMFORT-DISPLAY indicates the cause for the warning in form of a code
and an acoustic signal sounds in regular intervals.
These messages appear every 10 seconds.

Press enter

to quit the audible signal.

ALARM
CODE 01

Low liquid level alarm in the internal reservoir
The temperature system is operated without bath fluid, or the liquid level is
insufficient. Replenish the bath tank with the bath fluid.
Tube breakage has occured (insufficient filling level due to excessive bath fluid
pumped out). Replace the tubing and replenish the bath tank with the bath fluid.
The float is defect (e. g., because damaged in transit). Repair by authorized JULABO
service personnel.

WARNING
CODE 03

Excess temperature warning

WARNING
CODE 04

Low temperature warning

ALARM
CODE 05

Cable of the working temperature sensor interrupted or short-circuited.

ALARM
CODE 06

Defect of the working or safety temperature sensor.
Working temperature and safety sensors report a temperature difference of more than
50 K.

ALARM
CODE 07


Other errors (I

2

C-BUS errors)

ALARM
CODE 12

Error in A/D converter.

ALARM
CODE 13

Sensor in the internal reservoir >RES< is defect

The safety temperature value lies below the working temperature setpoint. Set

the safety temperature to a higher value.

ALARM
CODE 14

Safety sensor defect

 The safety temperature value lies below the working temperature setpoint. Set

the safety temperature to a higher value.