Alarm code 13 – JULABO Magnum 91 Temperature System User Manual

Page 37

background image

Temperature System Magnum 91

37








-OFF-

ALARM
CODE 13

ALARM!
IntAct: xx.xx°C
ExtAct: --.--°C
RESERVOIR

Excess temperature protection in the reservoir >RES<

This supplementary safety installation supervises and controls the
temperature of the bath liquid in the internal reservoir.

 Press the key

to indicate the safety temperature value in line 3

>RES< on the VFD-Info-Display and using a screwdriver simultaneously
turn the setting screw to the desired value
(example: 100 °C).

Setting range: 10 °C to 220 °C

in 2 °C steps

The alarm is indicated by optical and audible signals (continuous tone).
The error messages opposite appear on the DIALOG-DISPLAY (LCD) and
VFD-Info-Display.


 Press

enter

to quit the audible signal.

 After eliminating the malfunction, turn the mains power switch off and

on again to cancel the alarm state.


Warning:

Recommendation for external, closed systems:

 Set the safety temperature >TANK< 15 °C above the working temperature setpoint.

 The safety temperature value >RES

the bath liquid used.

Important:

Check the safety installations at least twice a year! . See page 20