beautypg.com

JULABO Magnum 91 Temperature System User Manual

Page 49

background image

Temperature System Magnum 91

49

Int.Programmer
>Edit
Delete
Set Clock

>Edit Profile: 1
Step: 6
Setp: 100.00°C
Time[h.m] 01:05

Examples:

Profile No 1:

Step

Setpoint

Time

Edit

Compile profiles:

 A flashing segment indicates that a number needs to be entered.

Under submenu "Edit Profile" enter a profile number.
Six profiles may be stored (nos. 0 to 5).

 Then programme the desired values for each section.

Use the keypad to set section number, target temperature and time

period. Set each entry with enter

.


When the program is running, only sections having complete
information for target temperature and time period are considered.
It makes sense, to leave out section numbers in the profile, in order to
use them later for corrections in the profile.
Important:
If a time of 00:00 is set for a profile, the profile is continued with the
next section only after the setpoint temperature (±0.2 °C) is reached.

IN

Press

escape

ESC

to return to the previous menu level.

Example:
Step

(No.)

1 3 5 6 9 11 14

Setpoint (°C)

50 50 100

100 80 80 20

Time

(h.m) 00:08 00.02 00:10 01:05 00:10 00:05 00:30

20

40

60

80

100

120

140

160

180

Steps

1

3

5

6

9 11

14

10

20

30

40

50

60

120 min

Step

Edit Profile: 1
>Step: 3
Setp: 50:00°C
Time[h.m] 00:02

Display sections:

1. Use the cursor keys

to select the submenu "Step", enter the

desired number and press enter

.

or
2.
If the submenu „Step“ is selected, section by section can be shown in

ascending order by the P-key

.

Each previously set value is displayed.
Die eingestellten Werte werden jeweils angezeigt.