Remote control – JULABO Magnum 91 Temperature System User Manual

Page 64

background image

Remote control

64

11. Remote control

11.1. Setup for remote control

9

6

5

1

SERIAL

S xx.xx

P x.xx

R

 Check the interface parameters for both interfaces (on circulator and PC) and

make sure they match. (Serial interface see page 54)

 Select the "Configuration" submenu and select the option "Setpoint" to define the

interface (see page 41).

If the temperature system is put into remote control mode via the configuration
level, the display will read
"R -OFF-" = REMOTE STOP.

 Connect both units with an interface cable.

Like all parameters which can be entered through the keypad, interface parameters
are stored in memory even after the circulator is turned off.


Factory settings:

RS232

BAUDRATE

4800 bauds

PARITY even

parity

HANDSHAKE

Protocol

RTS/CTS

(hardware

handshake)

Data

bits

7

Stop

bit

1