beautypg.com

JULABO Magnum 91 Temperature System User Manual

Page 11

background image

Temperature System Magnum 91

11


Temperature System

Magnum 91

230 V / 3 Phases / 60 Hz

Working temperature range

°C

-91 ... +250

Temperature stability

°C

±0.01 ... ±0.05

Temperature selection:

digital

via keypad

indication on DIALOG-DISPLAY (LCD)

remote control via personal computer

indication on monitor

Temperature indication:

VFD COMFORT-DISPLAY
LCD DIALOG-DISPLAY

Resolution

°C

0.01

Liquid level indication and
pump pressure stage indication

VFD COMFORT-DISPLAY
(Bar graph in five grades)

Temperature control

ICC - Intelligent Cascade Control

Working temperature sensor

Pt 100

Safety temperature sensor

Pt 100

Absolute Temperature Calibration

(ATC

int)

°C

±3

(ATC

ext)

°C

±10

Heater wattage

kW

6.0

Cooling capacity

°C

200 100

│ 20 0 -20 -40 -60 -80

(Medium: Julabo Thermal HL45 │ Ethanol)

pumpe stage 5

kW

3.0 4.8

│ 4.6 4.6 4.5 4.3 2.0 0.5

Cooling compressor

2-stage

Refrigerant

R404A / R23

Cooling unit

water cooled

Quantity passing

l/min

12 (at 20 °C inlet temperature)

Cooling water connectors

G 3/4“ external thread

Pressure pump

1)

adjustable in five grades

Stage

1

2

3

4

5

Pressure

max. at 0 Liter bar

0.8

0.95

1.2

1.4

1.9

Flow rate

max. at 0 bar

l/min

30

35

41

45

50

Filling volume

heat exchanger

min.

process

volume

usable expansion volume

liters
liters
liters

11.0
21.5
13.0

Dimenssions (BxTxH)

cm

71x88x165

Weight kg

385

Ambient temperature range

°C

5 ... 35

Noise level, 1 m distance

dBA

72

Protection class according to IEC 60 529

IP 20

Mains power connection 0 /-10 %

V/ Hz

230 / 3 Phases / 60

Current input at 230 V max. / phase

A

35

1)

Pump data in relation to fluids with a specific density of 1 kg/dm

3

All measurements have been carried out at: rated voltage and frequency, ambient temperature: 20 °C
Technical changes without prior notification reserved.