JULABO Magnum 91 Temperature System User Manual
Page 53

Temperature System Magnum 91
53
 
 
>EPROG: Current
 L Value 0.0°C
 H Value 300.0°C
ExtSet 50.0°C
EPROG : 50.00°C
IntAct: 50.00°C
Power: 50 %
Control: intern
3x
Example:
 Set the external voltage or current source output for the equivalent of
50 °C temperature setpoint.
The value adjusted and set on the external programmer is 
displayed in line 4 of the DIALOG-DISPLAY (LCD) for control purposes 
(Example: ExtSet: 50.0 
C).
After returning the LCD display to standard display by pressing
ESC
the
temperature value adjusted and set on the external voltage or current 
source is displayed in line 1 
(Example: EPROG 50.00 
C).
 
 
 
 
 
 
20
40
60
80
100
200
300
250
150
°C
1 2 3 4 5 6 7 8 9 10 V
2 4 6 8 10
14
18 20 mA
A
B
 
 
 
 
 
EPROG : 152.00°C
IntAct: 50.00°C
Power: 100 %
Control: intern
>EPROG: Current
 L Value 20.0°C
 H Value 200.0°C
ExtSet 152.0°C
3x
This EPROG input enables the use of different voltage and current 
values as program parameters. 
 "L Value" - Setting the LOW value:
1) Adjust and set the lowest desired value on the voltage or current 
source resp. (Example A: 1 V). 
Wait appr. 30 seconds. 
2) Assign a lower temperature threshold value to this adjusted 
voltage/current value by pressing the appropriate buttons on the 
keypad of the instrument (Example A: 20 
C ) and set by pressing
enter
.
 "H Value" - Setting the HIGH value:
1) Adjust and set the highest desired value on the voltage or current 
source resp. (Example A: 10 V). 
Wait appr. 30 seconds. 
2) Assign an upper temperature threshold value to this adjusted 
voltage/current value by pressing the appropriate buttons on the 
keypad of the instrument 
(Example A: 200
C) and set by pressing enter
.
IN
Return to the standard display by pressing escape
ESC
.
Example B in the diagram serves to illustrate that the end point values are 
freely selectable. 
Example out of diagram A:
IN
Adjusting the voltage source for an output of 7.6 V!
Line 4 of the DIALOG-DISPLAY (LCD) shows the externally set setpoint 
value. The instrument calculates this value from the rise angle of the two 
predecided end points (in example A: 7.6 V correspond to an external 
setpoint temperature of 152.0 
C ).
After returning the LCD display to standard display by pressing escape
ESC
, this value is displayed in line 1
(Example: EPROG 152.00
C).
Notice:
If this adjustment is not correctly performed at two different points, the setpoint setting will be 
incorrect. 