Cooling water connection – JULABO Magnum 91 Temperature System User Manual

Page 13

background image

Temperature System Magnum 91

13

2.5. Cooling

water

connection

Cooling water pressure (IN / OUT )

max.

6 bar

Difference pressure (IN - OUT )

3.5 to 6 bar

Cooling water temperature

<20 °C


Recommended quality of cooling water:

pH – value

7.5 to 9.0

Sulfate [SO4 2- ]

< 100 ppm

Hydrocarbonate [HCO3- ] / Sulphate [SO4 2-]

> 1 ppm

Hardness [Ca2+, Mg2+] / [HCO3-]

> 0,5 dH

Alkalinity

60 ppm < [HCO3-] < 300 ppm

Conductivity

< 500 μs / cm

Chloride (CL-)

< 50 ppm

Phosphate (PO43-)

< 2 ppm

Ammonia (NH3)

< 0.5 ppm

Free Chlorine

< 0.5 ppm

Ferri Ions (Fe3+ )

< 0.5 ppm

Mangano Ions (Mn2+)

< 0.05 ppm

Carbon dioxide (CO2)

< 10 ppm

Hydrosulfide (H2S)

< 50 ppm

Content of oxygen

< 0.1 ppm

Algae growth

impermissible

Suspended solids

impermissible


Notice:

Danger of corrosion of heat exchanger due to unsuitable quality of cooling water.

Due to its high content of lime hart water is not suitable for cooling and causes calcination
of the heat exchanger.

Ferrous water or water containing ferrous particles will cause formation of rust even in
heat exchangers made of stainless steel.

Chlorous water will cause pitting corrosion in heat exchangers made of stainless steel.

Due to its corrosive characteristics distilled and deionized water is unsuitable and will
cause corrosion of the bath. .

Due to its corrosive characteristics sea water is not suitable.

Due to its microbiological (bacteria) components which settle in the heat exchanger
untreated and unpurified river water and water from cooling towers is unsuitable.

Avoid particulate matter in cooling water.

 Avoid

putrid

water.

Notice:

Cooling

water

circuit

Risk of oil leaking from the cooling circuit (compressor) of the recirculating cooler into the cooling
water in case of a fault in the circuit!

Observe the laws and regulations of the water distribution company valid in the
location where the unit is operated.