Temperature System Magnum 91

27

Filling process:










TT

HT

Temp.

FILL

1. First bath fluid is filled in until 3 or 4 segments appear on the level indication

(see –Filling volume- page 26).
The reservoir is filled.

Add the filling volume for the external system (32)!

2. Press the key

to start the automatic filling process.

3. The liquid level in the unit falls.
4. When marker „C“ is reached, the message CODE 40 (low liquid level warning)

appears.
Then fill in the filling volume required for the external system (32) via the opening
(5).

5. When marker „D“ is reached, the message ALARM CODE 01 (low liquid level

alarm) appears.

Filling volume with closed, external system:
Bath fluids change in volume over a temperature increase/decrease. If 1 segment
or 4 segments should appear depends on the application that is to be carried out.
For temperature applications to +230 °C fill up to marker "HT", for applications to
for example –95 °C fill up to marker "TT".
(Temp. FILL = approx. ambient temperature).