background image

Electrical connections

62


1

2

3

4

5

6

REG+E-PROG

Programmer input / temperature recorder output (3.1)

Analog inputs / outputs see page 51

Pin

Signal

1 Voltage output

Channel 1

0 ... 10 V

2 Voltage output

Channel 2

0 ... 10 V

3 GND for outputs

0 V

4 Programmer input

EPROG

0 to 10 V / 0 to 20 mA

5 Current output

Channel 3

0 to 20 mA / 4 to 20 mA

6 GND for Progammer

0 V




1

3

2

ALARM





Alarm output
(for external alarm signal)
This potential-free change-over contact is activated in case of an alarm when
pins 2 and 3 are connected.

Switching capacity

max. 30 W / 40 VA

Switching voltage

max. 125 V

/

Switching current

max. 1 A

4

1

3

2

ext. Pt100




1

4

2

3

Shield

Plug

Look on

soldering side.

Pt100


Socket for external Pt100 sensor

Pin assignment:

Pin

Signal

1

I+

2

U+

3

U-

4

I-

The shield of the connecting cable is electrically connected to the plug housing
and the sensor tube.