JULABO Magnum 91 Temperature System User Manual

Page 35

background image

Temperature System Magnum 91

35

7.3.2. Warning

functions

  

The high and low temperature warning functions accompany the working
temperature value. An audible signal sounds in intervals when the actual
temperature exceeds one of the set limits (patented).
The corresponding message appears in line 4 on the DIALOG-DISPLAY
(LCD).

Setp.: 120.00°C
>Overt.:130.00°C
Subtmp:-55.00°C
St.Pump: 2

>Overt: Setting the high temperature limit

Setting in line 2.

The value previously set appears on the DIALOG-DISPLAY (LCD).
A flashing segment indicates that a value needs to be entered.

Use the keypad to enter the new value

(examples: 130.00 °C; -13.00 °C).

Press enter

to store the selected value.

Setp.: 120.00°C
Overt.:130.00°C
>Subtmp:110.00°C
St.Pump: 2

>Subtmp:

Setting the low temperature limit

Setting in line 3.

The value previously set appears on the DIALOG-DISPLAY (LCD).
A flashing segment indicates that a value needs to be entered.

Use the keypad to enter the new value

(examples: 110.00 °C; -17.00 °C).

Press enter

to store the selected value.

Note:
The warning functions will only be triggered when the actual bath temperature, after start from
the "OFF" or "rOFF" mode, lies within the set limits for 3 seconds.

Example:

Setp: 120.00°C
IntAct 130.50°C
ExtAct: --.--°C
Control: Intern

Excess temp.

t

150

130

120
110

20

°C

The corresponding message appears in line 4 on the DIALOG-DISPLAY
(LCD).