JULABO Magnum 91 Temperature System User Manual
Page 20
![background image](/manuals/563267/20/background.png)
Safety notes for the user
20
Pay attention to the thermal expansion of bath oil during heating to avoid
overflowing of the fluid.
Prevent water from penetrating into the hot bath oil.
Do not drain the bath fluid while it is hot!
Check the temperature of the bath fluid prior to draining (by switching the unit on
for a short moment for example).
Use suitable connecting tubing.
Avoid sharp bends in the tubing, and maintain a sufficient distance from
surrounding walls.
Make sure that the tubing is securely attached.
Regularly check the tubing for material defects (e.g. for cracks).
Never operate damaged or leaking equipment.
Always turn off the unit and disconnect the mains cable from the power source
before performing any service or maintenance procedures, or before moving the
unit.
Always turn off the unit and disconnect the mains cable from the power source
before cleaning the unit.
Always empty the bath before moving the unit.
Transport the unit with care.
Sudden jolts or drops may cause damages in the interior of the unit.
Observe all warning labels.
Never remove warning labels.
Never operate equipment with damaged mains power cables.
Repairs are to be carried out only by qualified service personnel.
Some parts of the unit and the pump connectors may become extremely warm or cold
during continuous operation. Therefore, exercise particular caution when touching these
parts.
Notice:
Check the safety installations at least twice a year!
Excess temperature protection according to IEC 61010-2-010
With a screw driver turn back the adjustable excess temperature protection until the shut-
down point (actual temperature).
Low level protection
To check the function of the float of this unit it cannot be operated manually.
The 5-graded level display should therefore be observed whenver refilling.
If the bath liquid thickens or cracks, the instrument should be cleaned and checked by
qualified personnel.