Configuration – JULABO Magnum 91 Temperature System User Manual
Page 40
Menu functions
40
8.1. Configuration
>Configuration
Controlparam.
Profile Start
Int.Programmer
Inputs/Outputs
Limits
Interface
Temp.Sensor
Pump
Configuration
Identif. xxxxx
Setpoint xxxxx
Autostart xxx
Standby xxxx
Language xxxxxx
IdentNo x
By means of the configuration functions, operation of the instrument can
be optimized for the current application.
Press
enter
to select the configuration submenu.
Use the up/down cursor keys
to select the desired option.
A flashing line indicates that a value needs to be entered.
Press the P-key
to select the parameter and press enter
.
Press
escape
ESC
to return the previous menu level.
> Identification
When performing an identification for the controlled system (temperature
application system), the control parameters Xp, Tn and Tv will be
automatically determined and stored.
Possible parameters:
off - no identification.
The control parameters ascertained during the last identification are
used for control purposes.
once - single identification
The instrument performs a single identification of the controlled
system after start.
After the identification process the parameter is automatically set to
"off".
always - continual identification
The instrument performs an identification of the controlled system
whenever a new setpoint is to be reached.
NOTE: Use this setting only when the temperature application
system changes permanently.
Important:
For physical reasons, refrigeration units provide full cooling performance
only after some minutes. During identification this is given consideration
by waiting for 5 minutes after switch-on.