Menu functions

40

8.1. Configuration

>Configuration
Controlparam.
Profile Start
Int.Programmer
Inputs/Outputs
Limits
Interface
Temp.Sensor
Pump

Configuration
Identif. xxxxx
Setpoint xxxxx
Autostart xxx
Standby xxxx
Language xxxxxx
IdentNo x


By means of the configuration functions, operation of the instrument can
be optimized for the current application.

 Press

enter

to select the configuration submenu.

 Use the up/down cursor keys

to select the desired option.

A flashing line indicates that a value needs to be entered.

 Press the P-key

to select the parameter and press enter

.

 Press

escape

ESC

to return the previous menu level.



> Identification
When performing an identification for the controlled system (temperature
application system), the control parameters Xp, Tn and Tv will be
automatically determined and stored.

Possible parameters:

off - no identification.

The control parameters ascertained during the last identification are
used for control purposes.

once - single identification

The instrument performs a single identification of the controlled
system after start.
After the identification process the parameter is automatically set to
"off".

always - continual identification

The instrument performs an identification of the controlled system
whenever a new setpoint is to be reached.
NOTE: Use this setting only when the temperature application
system changes permanently.



Important:

For physical reasons, refrigeration units provide full cooling performance
only after some minutes. During identification this is given consideration
by waiting for 5 minutes after switch-on.