Direct setting of the working temperature, Settings in the set menu – JULABO Magnum 91 Temperature System User Manual
Page 34
Manual operation
34
7.2. Direct setting of the working temperature
Setp.: 20.00°C
IntAct 21.00°C
ExtAct: --.--°C
Control: Intern
This setting may be carried out with the temperature system being in
operating state Start or Stop!
The value previously set appears on the DIALOG-DISPLAY (LCD)
(example: 20.00 °C).
A flashing segment indicates that a value needs to be entered.
Use the keypad to enter the new value and press enter
to
store the selected value.
(examples: 120.00 °C; 228.50 °C; -15.00 °C).
7.3. Settings in the SET menu
>Setp.: 120.00°C
Overt.:255.00°C
Subtmp:-55.00°C
St.Pump: 2
Press the key
to call up the SET menu. It is then indicated on
the DIALOG-DISPLAY (LCD).
Set the following values:
in Line 1: the working temperature in °C
Setp.:
Line 2: the high temperature limit in °C
Overt.:
Line 3: the low temperature limit in °C
Subtmp:
Line 4: the pump pressure stage
St.Pump:
Enter the desired values via the keypad. Settings may be carried out
with the temperature system being in operating state Start or Stop!
Use the cursors
to swap the lines.
Press
escape
ESC
to quit the SET menu.
7.3.1.
Setting the working temperature
>Setp.: 200.00°C
Overt.: 255.00°C
Suptmp: -55.00°C
St.Pump: 2
>Setp.: 120.00°C
Overt.:255.00°C
Subtmp:-55.00°C
St.Pump: 2
>Setp.:
Setting the working temperature
Setting in line 1.
The value previously set appears on the DIALOG-DISPLAY (LCD)
(example: 100.00 °C).
A flashing segment indicates that a value needs to be entered.
Use the keypad to enter the new value
(examples: 120.00 °C; 228.50 °C; -15.00 °C).
Press enter
to store the selected value.