Limits, Interface – JULABO Magnum 91 Temperature System User Manual

Page 54

background image

Menu functions

54

8.6. Limits

Int.Programmer
Inputs/Outputs
Limits
Interface
Temp.Sensor

Limits
>IntMax 400.00°C
IntMin -99.99°C
Band 30K
HeatingMax 100%
CoolingMax 100%

The limits IntMax and IntMin are only valid under external control (see 7.5.

Internal / external control). They restrict the temperature of the

internal bath to the desired maximum/minimum, also if the controller would
require a higher/lower temperature for the external system. As a result the
external setpoint may thus not be reached.

When operating the temperature system under external control, band
limiting is active. The preset value determines the maximum temperature
difference between the internal bath and the external load. This
adjustment possibility prevents sensitive equipment and temperature
devices from damage.

Heating and cooling power of the instrument are adjustable.
100 % corresponds to the values in the technical specifications of the
equipment.

 Select the submenu "Limits" with enter

.

 Select the desired option with the up/down cursor keys

.

A flashing digit indicates that a value needs to be entered.
IntMax

see temperature range LHxx

IntMin

see temperature range LHxx

Band

0 to 200 °C

HeatingMax

0 to 100 % in steps of 1 %

CoolingMax

0 to 100 % in steps of 1 %

 Set the newly entered value by pressing enter

.

 Return to the previous menu level by pressing escape

ESC

.

8.7. Interface

Interface
>Type RS232
Baudrate 4800
Parity no
Handshake softw

Inputs/Outputs
Limits
>Interface
Temp.Sensor

Interface
>Type RS232
Baudrate 4800
Parity no
Handshake softw

Interface
>Type RS485
Baudrate 4800
Parity no
Address 23

The interface parameters are set by selecting the submenu "Interface" on
the temperature system. Normally, this is a one-time-only adjustment.

 Press

enter

to select the submenu "Interface".

 Select the desired option with the up/down cursor keys

.

Enter

the desired value for the flashing digit.
Type

RS232 / RS485

Baudrate

4800/9600

Parity none/even/odd
Handshake software handshake/hardware handshake

Address 0 to 127

 Press the P-key

to select the desired parameter and set with

enter

.

 Return to the previous menu level with escape

ESC

.