Operator responsibility – Safety recommendations

12

Electrical connections:

Computer interface

RS232 / RS485

Programmer input -100 °C to 400 °C

=

0 to 10 V or 0 to 20 mA or 4 to 20 mA

0 to 10 V (0 V = -100 °C, 10 V = 400 °C)

0 to 20 mA (0 mA = -100 °C, 20 mA = 400 °C)

4 to 20 mA (4 mA = -100 °C, 20 mA = 400 °C)

Stand-by input

External alarm device

4-0 V DC / max. 25 mA

External measurement and control sensor

Pt100, 4-lead technique


Safety installations and warning functions:

Excess temperature protection >TANK<

adjustable from 0 °C ... 320 °C

Excess temperature protection >RES<

adjustable from 0 °C ... 220 °C

Low level protection

float switch

Liquid level display

optical, 5-graded

Classification according to DIN 12876-1

class III

Supplementary safety installations

High temperature warning function

optical + audible (in intervals)

Low temperature warning function

optical + audible (in intervals)

Supervision of the working sensor

plausibility control

Reciprocal sensor monitoring between
working and safety sensors

difference >50 K

Alarm indication

optical + audible (permanent)


Environmental conditions according to IEC 61 010-1:

Use indoors only.

Altitude up to 2000 m - normal zero.

Ambient temperature: see Technical specifications

Humidity:

Max. relative humidity 80% for temperatures up to +31 °C,

linear decrease down to 50% relative humidity at a temperature of +40 °C

The unit corresponds to Class I

Overvoltage category

II

Pollution degree

2

Caution:

The unit is not suitable for use in explosive environment

EMC requirements according to EN 61326-1

This unit is an ISM device classified in Group 1 (using high frequency for internal purposes), Class A (industrial
and commercial range).