List of commands – JULABO Magnum 91 Temperature System User Manual
Page 66
Remote control
66
11.3. List of commands
When the RS485 interface is used, the instrument address stands in front of each command (Axxx_).
in commands: Asking for parameters or temperature values to be displayed.
Command
Parameter
Response of instrument
version
none
Number of software version (V X.xx)
status
none
Status message, error message (see page 69)
in_pv_00
none
Actual bath temperature.
in_pv_01
none
Heating power being used (%).
in_pv_02
none Temperature
value
registered
by the external Pt100 sensor.
in_pv_03
none
Temperature value registered by the safety sensor. >TANK<
in_pv_05
none
Pump pressure in bar.
in_sp_00
none Working
temperature
in_sp_03
none
High temperature warning limit
in_sp_04
none
Low temperature warning limit
.
in_sp_05
none
Setpoint temperature of the external programmer
(socket - REG+E-PROG) .
in_sp_07
none
Pump pressure stage
in_hil_00
none
Max. cooling power (%).
in_hil_01
none
Max. heating power (%).
in_mode_02
none Identification
type:
0 = no identification
1 = single identification
2 = continual identification
in_mode_03
none
Type of the programmer input:
0 = Voltage
0 V to 10 V
1 = Current
0 mA to 20 mA
in_mode_04
none
Internal/external temperature control:
0 = Temperature control with internal sensor.
1 = Temperature control with external Pt100 sensor.