Menu functions

56

8.9. Pump




Pump
>Pump Off
Mode: drain

Limits
Interface
Temp.Sensor
>Pump

In this submenu the circulation pump can be turned on separately, without
heating element or cooling unit.
The menu option „Modus“ refers to the filling valve and the ventilation
valve. Depending on the operating status (drain / fill / sys open / sys close)
the valves are switched off automatically.

 Press

enter

to select the submenu "Pump".

 Select the desired option with the up/down cursor keys

.

Enter the desired value for the flashing digit.
Pump

OFF/ON

Mode:

drain / fill / sys open / sys close


 Press the P-key

to select the desired parameter and set with

enter

.

 Return to the previous menu level with escape

ESC

.

Explication of terms:
>Mode: System closed<
stands for closed, external system.

(Magnum

principle page 18)


In this operating status the filling valve is opened and the ventilation valve is closed.
>Mode: Sys open< stands for open external system
(not recommended by JULABO)

>Mode: drain<. (Chapter „Draining“ see page 31)
In this operating status the filling valve and the ventilation valve are opened.
In the configuration of this unit the >Mode: fill< means the same as >Mode: drain<.