Pump – JULABO Magnum 91 Temperature System User Manual
Page 56

Menu functions
56
8.9. Pump
 
 
 
 
Pump
>Pump Off
 Mode: drain
 Limits
 Interface
 Temp.Sensor
>Pump
In this submenu the circulation pump can be turned on separately, without 
heating element or cooling unit. 
The menu option „Modus“ refers to the filling valve and the ventilation 
valve. Depending on the operating status (drain / fill / sys open / sys close) 
the valves are switched off automatically.  
 Press
enter
to select the submenu "Pump".
 Select the desired option with the up/down cursor keys
.
Enter the desired value for the flashing digit. 
Pump 
OFF/ON
Mode:
drain / fill / sys open / sys close
  
 
 Press the P-key
to select the desired parameter and set with
enter
.
 Return to the previous menu level with escape
ESC
.
Explication of terms: 
>Mode: System closed<stands for closed, external system. 
 
(Magnum
principle page 18)
 
In this operating status the filling valve is opened and the ventilation valve is closed. 
>Mode: Sys open< stands for open external system 
(not recommended by JULABO) 
 
>Mode: drain<. (Chapter „Draining“ see page 31) 
In this operating status the filling valve and the ventilation valve are opened. 
In the configuration of this unit the >Mode: fill< means the same as >Mode: drain<. 
 
 
 