beautypg.com

Status messages, Error messages – JULABO Magnum 91 Temperature System User Manual

Page 69

background image

Temperature System Magnum 91

69

11.4. Status messages

The instrument sends data (including error messages) only when the computer sends a query.

Status messages

Description

00 MANUAL STOP

Magnum in „OFF“ state.

01 MANUAL START

or
01 MANUAL START,DEGASING

Magnum in keypad control mode

or

Magnum in keypad control mode and degasifying active

02 REMOTE STOP

Magnum in „r OFF“ state.

03 REMOTE START

or
03 REMOTE START,DEGASING

Magnum in remote control mode.
Or

Magnum in remote control mode and degasifying active


11.5. Error messages

Error messages

Description

-01 NIVEAU LEVEL ALARM

Unterniveau-Alarm im internen Reservoir

-02 REFRIGERATOR ALARM

Defect temperature sensor in refrigeration circuit.

-03 EXCESS TEMPERATURE

WARNING

High temperature warning "

".

-04 LOW TEMPERATURE

WARNING

Low temperature warning "

".

-05 WORKING SENSOR ALARM

Working temperature sensor short-circuited or interrupted.

-06 SENSOR DIFFERENCE ALARM

Sensor difference alarm.
Working temperature and safety sensors report a temperature
difference of more than 50 K.

-07 I

2

C-BUS ERROR

Internal error when reading or writing the I

2

C bus.

-08 INVALID COMMAND

Invalid command.

-09 COMMAND NOT ALLOWED IN

CURRENT OPERATING MODE

Invalid command in current operating mode.

-10 VALUE TOO SMALL

Entered value too small.

-11 VALUE TOO LARGE

Entered value too large.

-12 TEMPERATURE

MEASUREMENT ALARM

Error in A/D converter.