Menu functions

52



0 V

10 V

0 mA

20 mA

4 mA

10 °C

210 °C

197 °C

202 °C

 5 °C

 200 °C



Examples:

lowest temperature value:

10 °C

highest temperature value

210 °C

Fig. shows 200

C scaled to paper width

rise: 50 mV/°C


lowest temperature value:

197 °C

highest temperature value:

202 °C

Fig. shows 5

C scaled to paper width

rise: 2000 mV/°C





1.

REG+E-PROG

Configuration
Identif. xxxxx
Setpoint EPROG
Autostart xxx

Inputs/Outputs
chan.2: ActInt
chan.3: ActExt
>EPROG: Current

>EPROG: Current
L Value 20.0°C
H Value 300.0°C
ExtSet 50.0°C

2.

3.


EPROG - Input

This input is necessary when the nominal value is to be determined and
set by an external programmer (see page 41 )

1. Connect the external voltage or current source or programmer to

socket (3.1) REG+E-PROG of the instrument.

2. Adjust the menu option >Setpoint< on >EPROG< in the menu

>Konfiguration< (see page 41)

3. The programmer (E-PROG) input of the instrument can be matched to

the output signal of the external voltage or current source.

Voltage

voltage

input

Current

current

input

Select the desired input value with the P-key

and set with enter

.

"L Value" - Setting the LOW value::

First adjust and set the lowest voltage or current on the external
voltage or current source (e.g. 0 V or 0 mA).
Then after approx. 30 secs enter the corresponding temperature value
(e.g. 20.00°C).on the instrument by pressing the appropriate buttons

on the keypad and press enter

to set.

"H Value" - Setting the HIGH value:

First adjust and set the highest voltage or current on the external
voltage or current source (e.g. 10 V or 20 mA).
Then after approx. 30 secs enter the corresponding temperature value
(e.g. 300 °C).on the instrument by pressing the appropriate buttons on

the keypad and press enter

to set.

 Return to the standard display by pressing escape

ESC

.