JULABO Magnum 91 Temperature System User Manual
Page 52
Menu functions
52
0 V
10 V
0 mA
20 mA
4 mA
10 °C
210 °C
197 °C
202 °C
5 °C
200 °C
Examples:
lowest temperature value:
10 °C
highest temperature value
210 °C
Fig. shows 200
C scaled to paper width
rise: 50 mV/°C
lowest temperature value:
197 °C
highest temperature value:
202 °C
Fig. shows 5
C scaled to paper width
rise: 2000 mV/°C
1.
REG+E-PROG
Configuration
Identif. xxxxx
Setpoint EPROG
Autostart xxx
Inputs/Outputs
chan.2: ActInt
chan.3: ActExt
>EPROG: Current
>EPROG: Current
L Value 20.0°C
H Value 300.0°C
ExtSet 50.0°C
2.
3.
EPROG - Input
This input is necessary when the nominal value is to be determined and
set by an external programmer (see page 41 )
1. Connect the external voltage or current source or programmer to
socket (3.1) REG+E-PROG of the instrument.
2. Adjust the menu option >Setpoint< on >EPROG< in the menu
>Konfiguration< (see page 41)
3. The programmer (E-PROG) input of the instrument can be matched to
the output signal of the external voltage or current source.
Voltage
voltage
input
Current
current
input
Select the desired input value with the P-key
and set with enter
.
"L Value" - Setting the LOW value::
First adjust and set the lowest voltage or current on the external
voltage or current source (e.g. 0 V or 0 mA).
Then after approx. 30 secs enter the corresponding temperature value
(e.g. 20.00°C).on the instrument by pressing the appropriate buttons
on the keypad and press enter
to set.
"H Value" - Setting the HIGH value:
First adjust and set the highest voltage or current on the external
voltage or current source (e.g. 10 V or 20 mA).
Then after approx. 30 secs enter the corresponding temperature value
(e.g. 300 °C).on the instrument by pressing the appropriate buttons on
the keypad and press enter
to set.
Return to the standard display by pressing escape
ESC
.