Draining, S 20.00 p x.xx – JULABO Magnum 91 Temperature System User Manual
Page 31
Temperature System Magnum 91
31
5.8. Draining
Notice:
Do not drain the bath fluid while it is hot!
Check the temperature of the bath fluid prior to draining (by switching the unit on for a
short moment, for example).
Store and dispose the used bath liquid according to the laws for environmental
protection.
Please observe the regulations for disposal which are effective in the country of
operation.
-OFF-
S 20.00
P x.xx
Pump
>Pump OFF
Modu: drain
Turn on the temperature system with the mains power switch or
press the stop key
to make the unit enter the STOP-
MODE.
In the submenu >Pump<, set the menu items >Mode< to
>drain<.
(see page 29 Adjusting Mode:)
Place a suitable vessel for accepting the used bath liquid
underneath the drain.
Slide a short piece of tubing onto the drain port and hold the end
into the vessel..
Unscrew the drain screw by some turns.
The unit has two drain screws. Both have to be used.
The liquid level in the unit falls.
Now the VFD-Display of the unit should show low level,
WARNING >CODE 40< and then ALARM >CODE 01<.
Press enter
to quit the audible signal.
Drain the external system:
At the upper nozzle of the external system, disconnect the
tubing to the pump connector "return" (12) until an air
stream moves in.
Tighten the drain screw after draining the unit.
(Cleaning the unit see page 73)