beautypg.com

9 technical data – HYDAC CMU 1000 User Manual

Page 117

background image

Condition Monitoring Unit CMU 1000

Page 117

Revised 17.12.2009

HYDAC ELECTRONIC GMBH Mat.-No.:

669749

9 Technical Data

9.1 Power Supply

- Input voltage:

18.0 ... 35.0 V DC

- Power consumption:

max. 1.5 A (3.5 A with connected CSI-F-10)

- Reverse voltage protection:

-30 V

- Dielectric strength:

+40 V


9.2 Connecting Sensors

Up to 8 sensors with HSI functionality or
up to 8 SMART sensors and in addition up to 8 analog sensors, as well as
up to 4 digital sensors

(4 x digital / 2 x digital + 2 x frequency / 3 x digital + 1 x frequency)

can be connected at the same time


The measuring ranges of the analog inputs are designed for a variety of requirements


9.3 Analog Inputs

Channels I and J (precision)
4 .. 20 mA

(

≤ ± 0.1 % FS max.)

0 .. 20 mA

(

≤ ± 0.1 % FS max.)

0.5 .. 4.5 V

(

≤ ± 0.1 % FS max.)

0 .. 10 V

(

≤ ± 0.1 % FS max.)



Channels M and N (precision)
4 .. 20 mA

(

≤ ± 0.1 % FS max.)

0 .. 20 mA

(

≤ ± 0.1 % FS max.)

0.5 .. 4.5 V

(

≤ ± 0.1 % FS max.)

Channels K and L (precision)
4 .. 20 mA

(

≤ ± 0.1 % FS max.)

0 .. 20 mA

(

≤ ± 0.1 % FS max.)

0.5 .. 4.5 V

(

≤ ± 0.1 % FS max.)

0 .. 50 V

(

≤ ± 0.1 % FS max.)

-10 .. +10 V (

≤ ± 0.2 % FS max.) L only!


Channels O and P (precision)
4 .. 20 mA

(

≤ ± 0.1 % FS max.)

0 .. 20 mA

(

≤ ± 0.1 % FS max.)

0.5 .. 4.5 V

(

≤ ± 0.1 % FS max.)

-10 .. +10 V (

≤ ± 0.2 % FS max.) P only!

9.4 Digital Inputs

- Quantity: 4, of which 2 are for frequency measurements (Channels Q and R)
- Trigger threshold:

approx. 2 V

- Dynamics:

30 kHz


9.5 Measurement Channels

A total of up to 32 measurement channels can be processed by the CMU.
One measurement channel can be a value of a connected sensor (also a subchannel
or a SMART sensor) or a value derived (calculated) from sensor data.