Det-Tronics PIRECL Infrared Hydrocarbon Gas Detector PointWatch Eclipse User Manual
Page 46
F-7
95-8526
14.2
5 history
This menu option shows extensive historical information about the detector. Data available includes:
1 Running hrs xxxx (the number of hours the unit has been powered).
2 Max temperatures (the maximum temperatures recorded in the device).
See max temperature submenu below.
3 Min temperatures (the minimum temperatures recorded in the device).
See min temperature submenu below.
4 Cal log (data regarding stored calibrations). The most recent calibration is shown first.
Calibrations are recorded as zero only cal, cal OK (zero and span were successfully completed),
and cal failed. See cal log submenu below.
5 Event log (data regarding stored events). The most recent event is shown first.
Recorded events include blocked optics, warm-up, zero drift, low alarms and high alarms.
See event log submenu below.
Max temperature submenu:
Running hrs xxxx
Maximum Temperature
xxxx degC xxxx hours
Max temp since reset
xxxx degC xxxx hours
Reset min&max temp?
ABORT OK
Min temperature submenu:
Running hrs xxxx
Minimum Temperature
xxxx degC xxxx hours
Min temp since reset
xxxx degC xxxx hours
Reset min&max temp?
ABORT OK
cal log submenu:
Running hrs xxxx
Calibration history
(Event)
xxxxx Hrs
1 Previous
2 Next
3 Exit
event log submenu:
Running hrs xxxx
Event history
(Event)
xxxx Hrs
1 Previous
2 Next
3 Exit