beautypg.com

Det-Tronics PIRECL Infrared Hydrocarbon Gas Detector PointWatch Eclipse User Manual

Page 18

background image

16

95-8526

14.2

reMote Calibration option
In most applications, it is recommended to install the

Pointwatch Eclipse where it will contact the vapor of

interest as quickly as possible. Unfortunately, the best

location for early warning can often result in accessibility

problems for the operator when calibration is required.

In these applications, the Model PIRTB Termination Box

is highly recommended to provide the ability to calibrate

the Pointwatch Eclipse from a remote location.

The PIRTB consists of a termination/circuit board, housed

within an explosion-proof junction box. The circuit board

contains a magnetic reed switch for initiating calibration,

an indicating LED to signal the operator when to apply

and remove the calibration gas, and a wiring terminal

block. The junction box cover is furnished with a small

viewing window that enables calibration to be performed

without hazardous area de-classification. The PIRTB may

be installed up to 100 feet away from the Eclipse. Refer

to Figure 21 for remote calibration configuration options.

NOTE

The remote calibration switch is intended for
initiating calibration only. Resetting latching alarm
outputs using the remote calibration switch cannot
be accomplished without entering the Calibration
mode.

The following recommendations are provided to enhance

operator ease and convenience of remote calibration

configurations:

1. Install the Eclipse in such a manner that the onboard

LED is visible whenever possible. This will aid in

checking device status "at a glance."

2. The Eclipse is provided with a calibration gas nozzle

on the weather baffle, which allows the use of

permanently attached calibration gas delivery tubing

(either polyethylene or stainless steel). The tubing is

typically routed in parallel with the remote calibration

cabling to the same location as the PIRTB termination

box. This arrangement enables a technician to

initiate calibration and deliver the calibration gas to

the Eclipse from a single location.

3. When permanently installed calibration gas tubing

is utilized, always install a shut-off valve at the open

end to prevent unwanted vapors or debris from

entering the tubing.

4. Always purge the permanent tubing with clean,

dry compressed air prior to and immediately after

calibration to ensure that residual combustible

gases are cleared. Always close the shutoff valve

after post-calibration purging is complete. This will

ensure that all hydrocarbon vapors are eliminated

from the Eclipse optics.

5. Note that permanently installed calibration gas tubing

will increase the calibration gas consumption rate as

a function of total tubing length.

Other methods of remote Eclipse calibration include

utilizing HART or MODBUS communications. Refer to

the HART and MODBUS appendices for details.

SpeCial appliCationS
The standard Pointwatch Eclipse is intended for

open area combustible gas detection applications.

However, special detector configurations are available

for applications such as duct-mounting and sample

extraction. Contact Detector Electronics Corporation for

information on these special device configurations.