Det-Tronics PIRECL Infrared Hydrocarbon Gas Detector PointWatch Eclipse User Manual
Page 23

21
95-8526
14.2
calibRation
Calibration oVerView
Although routine calibration of the Pointwatch Eclipse is
normally not required, the device supports non-intrusive
field calibration capability. Two (2) calibration procedure
options are provided:
1. normal Calibration is a two-step process consisting
of clean air (zero) condition and mid-scale (span)
adjustment. Calibration gas must be applied by
the operator to enable span adjustment. Normal
calibration is required whenever the gas type setting
has been changed from the factory setting (methane,
propane, ethylene, or butane). Purge the Eclipse
optics with clean, dry air prior to calibration initiation
to ensure that an accurate zero (clean air) condition
is present.
The following Normal Calibration guidelines always
apply:
A. The Eclipse is factory set for detection of
methane, propane, ethylene, or butane. If the
gas setting is changed (using HART, MODBUS
or EQP communication), the Eclipse must be
re-calibrated (normally with the matching gas
type).
B. The calibration gas type normally should match
the selected gas setting for the Eclipse. Different
calibration gas types are selectable using HART,
MODBUS or EQP communication.
C. The recommended calibration gas concentration
is 50% LFL, although other calibration
concentrations may be utilized if previously
defined in the Eclipse using HART, MODBUS or
EQP communication.
2. zero only Calibration is a one-step process
consisting of clean air (zero) condition adjustment
only, which is performed automatically by the device.
This procedure adjusts the "clean air" signal output
only, and is normally used if the 4 milliampere signal
level has drifted. The cause of drift is typically due to
the presence of background gas during calibration.
Purge the Eclipse optics with clean, dry compressed
air prior to calibration initiation to ensure an accurate
zero (clean air) condition is present.
aDDitional Calibration noteS
Important
Always ensure that the correct gas type is used for
calibration. (2.5 LPM flow rate is recommended.)
NOTE
It is highly recommended that the detector has
been operating for at least two hours before
calibrating.
NOTE
Always ensure that the Eclipse optics are totally
free of all hydrocarbons before initiating calibration.
This may require purging of the Eclipse with pure
air prior to initiating calibration.
NOTE
Under very windy conditions, it may not be
possible to successfully calibrate the Eclipse. This
situation is easily corrected by using the Eclipse
Calibration Bag (P/N 006682-002), available from
Det-Tronics.
NOTE
Always place the protective cap back on
the calibration nozzle after completion of span
calibration.