WIKA CPC6000 User Manual
Page 91

Automated Pressure Calibrator
CPC 6000
Mensor/WIKA Operating Instructions - CPC 6000
91
To get to the Calibration screen press [Setup], [Sensor], and then press [Calibrate].
10.6 Password
Autozero does not require a password, however one is needed to change any date of calibration, Zero,
Span, or to change the system password. The password installed at the factory was 1 2 3 4 5 6, but
the user can change this as described below.
10.6.1 Change Password
To change the current password:
1.
Press the [Calibrate] key to see the Calibration Data screen similar to Figure 10.6.1A. Notice the
padlock.
Figure 10.6.1A - Calibration Data Screen (locked)
2.
Touch the [Unlock] key to display the Enter Password window as shown in Figure 10.6.1B.
Figure 10.6.1B - ‘Enter Password’ Window
- 890.09.2190 (44 pages)
- A-10 (96 pages)
- A2G-50 (52 pages)
- A2G-55 (36 pages)
- AC-1 (88 pages)
- C-2 (24 pages)
- D-10-7 (112 pages)
- D-20-9 (51 pages)
- D-20-9 (35 pages)
- DG-10 (112 pages)
- DP-10 (44 pages)
- DPT-10 (96 pages)
- DPT-10 (92 pages)
- GCS-1 (76 pages)
- HP-2 (84 pages)
- IL-10 (31 pages)
- IPT-10 (56 pages)
- IPT-10 (48 pages)
- IPT-10 (28 pages)
- IPT-10 (36 pages)
- IS-20-F (7 pages)
- IS-20-F (43 pages)
- IS-20-H (7 pages)
- UT-10 (42 pages)
- UT-10 (52 pages)
- IUT-10 (78 pages)
- LH-10 (60 pages)
- LH-20 (60 pages)
- LS-10 (60 pages)
- MG-1 (92 pages)
- MH-1 (11 pages)
- MH-2 (9 pages)
- MHC-1 (84 pages)
- N-10 (35 pages)
- O-10 (108 pages)
- OT-1 (60 pages)
- P-30 (92 pages)
- PSA-31 (124 pages)
- PSD-30 (128 pages)
- R-1 (92 pages)
- S-10 (35 pages)
- S-11 (102 pages)
- S-20 (52 pages)
- S-20 (96 pages)
- SL-1 (21 pages)