WIKA CPC6000 User Manual
Page 60
Automated Pressure Calibrator
CPC 6000
60
Mensor/WIKA Operating Instructions - CPC 6000
7.4.7 Remote Emulation
The Mensor PCS 400 is an earlier generation instrument similar to the CPC 6000. There is some
compatibility between the CPC 6000 and a PCS 400 in that the CPC 6000 will respond to many of the
remote instructions as if it were the older instrument. The PCS 400 commands will operate only on the
currently active control channel.
Table 7.4.7.1 is a list of the remote commands and queries which the CPC 6000 will recognize and
respond to.
A space between elements in a command indicates a required delimiter. Use a space where such a
delimiter is indicated. A full description of the syntax and use for each of these commands and queries
are presented in the PCS 400 manual.
i
Notice
All pressure values will be in the currently active pressure units
unless otherwise stated.
7.4.7.1 PCS 400 Emulated Commands and Queries
Table 7.4.7.1 - PCS 400 Emulated Commands and Queries
Ref
Command
Data
Function / Response
1
_pcs4 autorange
0 or 1
1 turns autorange on, 0 turns autorange off.
2
_pcs4 autorange?
Returns 1 if in autorange, 0 if in range hold.
3
_pcs4 cal span
Sets the span of the active transducer to
4
_pcs4 cal zero
Sets the zero of the active transducer to
5
_pcs4 cal_disable_off
Enables zero or span calibrations if previously disabled.
6
_pcs4 cal_disable_on
Prevents zero or span calibrations.
7
_pcs4 ctrl
Sets control value. Will take effect immediately if instru-
ment is in control mode.
8
_pcs4 ctrl?
Returns the current control point in current engineering
units.
9
_pcs4 ctrlmax
Sets maximum control value.
10
_pcs4 ctrlmax?
Returns current maximum control pressure.
11
_pcs4 ctrlmin
Sets minimum control value.
12
_pcs4 ctrlmin?
Returns current minimum control pressure.
13
_pcs4 default
Sets default values into instrument.
14
_pcs4 err?
Returns the error number and description.
15
_pcs4 func ctrl
Instrument placed in control mode at
in
16
_pcs4 func F1
Toggles between absolute and gauge modes if an inter-
nal barometric sensor is installed.
17
_pcs4 func meas
Instrument placed in measure mode.