WIKA CPC6000 User Manual
Page 58

Automated Pressure Calibrator
CPC 6000
58
Mensor/WIKA Operating Instructions - CPC 6000
7.4.6 CPC 6000 Error Codes
Table 7.4.6 - CPC 6000 Error Codes
Code
Serial
Poll
Byte
Description
Error String Returned
E00
00h
reserved
unused code
E01
41h
GPIB listen/talk error
GPIB LISTEN/TALK ERROR
E02
42h
Syntax error
UNKNOWN COMMAND
E03
43h
Expected a valid _PCS4 command
EXPECTED A VALID _PCS4 COMMAND
E04
44h
Expected a valid FUNC command
EXPECTED A VALID FUNC COMMAND
E05
45h
Expected a valid CAL command
EXPECTED A VALID CAL COMMAND
E06
46h
Expected a valid TEST command
EXPECTED A VALID TEST COMMAND
E07
47h
Expected a pressure units selection
EXPECTED A PRESSURE UNITS SELEC-
TION OR INVALID TERMINATION STRING
E08
48h
Expected a pressure value
EXPECTED A PRESSURE VALUE
E09
49h
Expected a transducer selection
EXPECTED A XDUCER SELECTION
E10
4ah
Expected a seconds selection
EXPECTED A SECONDS SELECTION
E11
4bh
reserved
unused code
E12
4ch
reserved
unused code
E13
4dh
Invalid pressure units selection
INVALID PRESSURE UNITS SELECTION
E14
4eh
Invalid control pressure value selection INVALID CONTROL PRESSURE VALUE
SELECTION
E15
4fh
Invalid rate value selection
INVALID RATE VALUE SELECTION
E16
50h
Invalid a/d unit selection
INVALID A/D UNIT SELECTION
E17
51h
Invalid zero offset value selection
INVALID ZERO OFFSET VALUE SELECTION
E18
52h
Invalid Span offset value selection
INVALID SPAN OFFSET VALUE SELECTION
E19
53h
Invalid rate units selection
INVALID RATE UNITS SELECTION
E20
54h
Sensor overrange
SENSOR OVERRANGE
E21
55h
Sensor underrange
SENSOR UNDERRANGE
E22
56h
Sensor failure detected
SENSOR FAILURE DETECTED
E23
57h
Low source pressure
LOW SOURCE PRESSURE
E24
58h
Regulator failure detected
REGULATOR FAILURE DETECTED
E25
59h
reserved
unused code
E26
5ah
Internal leak detected
INTERNAL LEAK DETECTED
E27
5bh
Program error detected
PROGRAM ERROR DETECTED
E28
5ch
Memory error detected
MEMORY/COEFFICIENT ERROR DETECT-
ED
E29
5dh
External Leak detected
EXTERNAL LEAK DETECTED
E30
5eh
High exhaust pressure
VACUUM ERROR DETECTED
E31
5fh
Transducer error detected
XDUCER ERROR DETECTED