WIKA CPC6000 User Manual
Page 53

Automated Pressure Calibrator
CPC 6000
Mensor/WIKA Operating Instructions - CPC 6000
53
39
Errorno?
Returns CPC 6000 error code and text.
40
Filter
{Off, Low, Normal, High}
Sets the reading filter.
41
Filter?
Returns the reading filter.
42
Gasdensity
{value}
Sets the head pressure gas density in lb/ft^3.
43
Gasdensity?
Returns the head pressure gas density in lb/ft^3.
44
Gastemp
{value}
Sets the head pressure gas temperature in °F.
45
Gastemp?
Returns the head pressure gas temperature.
46
Gateway
{string e.g. nnn.nnn.nnn.nnn}
Sets the Ethernet gateway address.
47
Gateway?
Returns the Ethernet gateway address.
48
Height
{value}
Sets the head pressure height in inches.
49
Height?
Returns the head pressure height.
50
Highspeed
{Yes or No}
Sets the active channel to high speed mode:
.006% stablewindow and 1 second stabledelay.
51
ID?
v.vv}
Returns the instrument identity where {ssssss}
is the serial number, and {v.vv} is the software
version number.
52
IP
{string e.g. nnn.nnn.nnn.nnn}
Sets the IP address of the instrument.
53
IP?
Returns the IP address of the instrument.
54
Keylock
{Yes or No}
YES to lock, or NO to unlock the on-screen
keys.
55
Keylock?
Returns current keylock status as YES or NO.
56
List?
Returns list of available turndowns on installed
sensors in the active channel. X will be non-ex-
istent if the turndown is not available.
57
Listcal?
dd/yyyy},{td},{mm/dd/
yyyy};SEC,{td},{mm/dd/
yyyy},{td},{mm/dd/yyyy}
Returns the serial number of each installed
transducer and calibration dates for each range.
58
Listrange?
{min},{max};SEC,{td},{min},{m
ax},{td},{min},{max}
Returns the minimum and maximum ranges of
all installed sensors for the active channel.